Tympanogram Interpretation
Audiograms
To begin testing, the probe, fitted with an appropriately sized eartip, is
positioned at the opening of the ear canal. For operator convenience, the probes
of the Auto Tymps are equipped with either signal lights (GSI 38) or an easy to
read display (GSI 37) to keep the operator informed of the test status at all
times. Even inexperienced instrument users can obtain reliable results with
confidence. When a tight seal is
obtained, a known quantity of sound energy is introduced to the ear. By
introducing a known amount of sound energy to the ear, a measurement of the
energy not transmitted into the middle-ear system is recorded. The amount
of sound energy transmitted is equal to the amount of sound energy introduced,
minus the amount of sound energy that returns to the probe microphone.
The amount of energy transmitted is
directly related to the compliance of the system. Compliance (labeled in cm3 of
equivalent volume) indicates the amount of mobility in the middle ear. For
example, the more energy returned to the probe, the less energy admitted, and
thus a low compliance measurement. Low compliance measurements indicate a stiff
or obstructed middle ear (See Figure 3). Of course, the opposite is true
for a high amount of energy transmission, indicating a flaccid or highly mobile
system.
At this point, the probe introduces a
pressure of + 200 daPa
to the middle ear canal. Decapascals, or daPa is a measurement of air
pressure; whereby, 1.02mm H20 = 1.00 daPa. This positive pressure
forces the tympanic membrane inward and
the approximate ear canal volume is recorded. This volume gives a
baseline from which the compliance curve is drawn. The
pressure is now varied in the negative
direction, while constantly monitoring the compliance of the system. The
pressure continues toward the negative direction until a pressure peak has been
detected and -100 daPa has been reached or until a pressure of -400 daPa is
present in the ear canal, whichever comes first.
Once the pressure is equalized on both
sides of the tympanic membrane, the point of peak compliance is stored, as is
the pressure at which peak compliance occurs.
A tympanogram is a graphic representation of the
change in compliance (Y or
vertical axis) of the middle-ear system as air
pressure (X or horizontal axis)
is varied. Valuable diagnostic information is obtained from both the graphic and
quantitative data (See Results Section).
1. Ear Canal Volume (0.2-2.0), measured in cm3, indicates the volume from
the probe tip to the tympanic membrane at a pressure of +200 daPa. Ear canal
volume < 0.2: cerumen or probe malposition. Ear canal volume > 2: PE tubes or
perforation.
2. Compliance Peak (0.2-1.4), expressed in daPa, indicates the amplitude
of the peak. This value can vary from NP (no peak,
when compliance < 0.1) to 6.0 cm3. If <0.2, 'stiff'; if > 1.4, 'hyperflaccid.'
3. Pressure Peak (-150 to 100), measured in cm3, indicates the pressure
at which equalization occurs on both sides of the tympanic membrane. It also
indicates the pressure at which peak compliance or maximum mobility is attained.
This corresponds to the value on the horizontal axis of the graph.
Pressure peaks more negative than –150 indicative of poorly
functioningeustachiantube or possibly some fluid. Extreme positive pressures
possibly at the onset of AOM
4. The Scale of Reference, measured in cm3, is dependent on the amplitude
of the tymp peak measurement. This scale is either 1.5 cm3 or 3.0 cm3. Should
the peak measure 1.5 or less in amplitude, the scale reading will be 1.5 cm3. If
it is 1.6 or greater, the scale will read 3.0 cm3. The change in scale size
merely allows a greater distribution of the graph on the chart. A tymp peak
reading of "NP" will automatically cause the scale reading to be 1.5, which is
indicative of no peak.
5. Normal Box indicates the range of pressure peak and compliance peak
values associated with normal middle-ear function. (-150 daPa to +100 daPa, 0.2
cm3 to 1.4 cm3 per ASHA, 32, Suppl. 2, 1990, 17-24).
6. Gradient (60-150 child, 50 to 110 adult), expressed in daPa, is the
tympanometric pressure width at 50% of the compliance peak. The width of the
pressure curve; the range of pressures needed for mobility Infants may show
higher gradient values due to the mobility of their ear canals. Wide gradients
indicate presence of fluid. Narrow gradients associated with a flaccid middle
ear system. Near zero gradient: viscous effusion.
7. Acoustic Reflex, expressed as a yes (normal)/no, dB HL or dB HL and a
curve, signifies the level at which the acoustic reflex causes the stapedial
muscle to contract. This value will depend upon the level or intensity at which
the acoustic reflex is detected at a given frequency. "NR" indicates no reflex
occurred at the level tested. "NT" indicates test aborted prior to completion.
(negative if hearing loss is present)
In assessing the results of tympanometry, they should be analyzed with respect
to:
(1) Ear Canal Volume. (2) Compliance (peak amplitude)
(3) Gradient (tymp width) (4) Middle-Ear Pressure (peak pressure)
Each result is a piece of the puzzle and provides information on a possible
cause for any abnormality. The gradient
acts as an adjunct to the tymp
and ear canal volume measurements by helping to differentiate between
tymps
with similar peak values. A larger than-normal gradient value can indicate the
presence of fluid in the middle ear when other parameters are within normal
limits.
The configuration of the tympanogram itself indicates the compliance of the
middle ear system. In the physiologically stiffened ear, the variation in
mobility will be less, indicating a limited movement at the point of equalized
pressure. A hyperflaccid condition, indicative of a scarred membrane is
illustrated by a sharp peak.
Tympanograms can be characterized by their shape and location. The
key characteristics affecting the
pathology diagnoses are; pressure peak, compliance (maximum displacement) and
width (gradient) of tympanograms.
Cases of otitis
media can easily be detected by a depressed compliance peak amplitude and a
negative pressure peak. (In general, the shape of the peak is more
sensitive/specific vs
peak pressure or peak compliance - the more rounded the peak, or ultimately an
absent peak, the higher the probability that an effusion is present. Nelson 16h
ed) The results provide an objective and tangible means for a simple
diagnosis. The example tymps that follow are analyzed and the results could be
interpreted as stated below each chart.


Type A curve
Normal Ear Canal Volume
Normal Middle-Ear Mobility
Normal Middle-ear Pressure (o)
Normal GR

Poorly Functioning Eustachian Tube
Type C Curve
Normal Ear Canal Volume
Restricted Mobility (low amplitude)
Abnormal Middle-ear Pressure (very negative)
Borderline wide GR
POSSIBLE CAUSE
Poorly functioning Eustachian Tube
Possibly Some Fluid

Partially Blocked Eustachian Tube
Normal Ear Canal Volume
Normal Middle-ear Mobility
Slightly Negative Middle-ear Pressure
Normal GR
POSSIBLE CAUSE
Partially Blocked Eustachian Tube

Possible Nasal Congestion
Normal Ear Canal Volume
Normal Middle-ear Mobility
Positive Middle-ear Pressure
Normal GR
POSSIBLE CAUSE
Patient has a cold

Type As curve
Normal Ear Canal Volume
Restricted Mobility
Normal Middle-ear Pressure
Borderline-or-Wide GR
POSSIBLE CAUSE
Glue-ear
Otosclerosis; stiff ossicular chain
Severely Scarred Tympanic Membrane
Plaque over Tympanic Membrane

Serous Otitis Media with Air Pockets
(Serous OM without airpockets would show a flat line)
Type B curve
Normal Ear Canal Volume
Restricted Mobility
Negative Middle-ear Pressure
Abnormally Wide GR
POSSIBLE CAUSE
Serous Otitis Media
Small Air Pockets Present

Ossicular Disarticulation/Discontinuity
Type AD curve
Normal Ear Canal Volume
Extremely Hyperflaccid Middle-ear
Normal Middle-ear Pressure
Narrow GR
POSSIBLE CAUSE
Ossicular Disarticulation
More Detailed Testing Indicated

Scarred Tympanic Membrane
Normal Ear Canal Volume
Hyperflaccid Middle-ear System
Normal Middle-ear Pressure
Normal-or-Borderline Narrow GR
POSSIBLE CAUSE
Minor Scar Tissue in Tympanic Membrane

Tympanic Membrane Perforation
Abnormal Ear Canal Volume (large)
No Mobility
No Middle-ear Pressure
No GR
POSSIBLE CAUSE
Open Perforation
Patent Pressure Equalization (P-E) Tube

Serous Otitis Media
Normal Ear Canal Volume
No Mobility
No Middle-ear Pressure
No GR
POSSIBLE CAUSE
Fluid-filled Middle-ear (Serous Otitis Media)
Compliance Peak May be Present at a Much More Negative Pressure Than -400daPa

Effusion
Reference
http://www.grason-stadler.com/tymp.html